Incidence of delivery wire recapture failure with the Pipeline Flex device
Document Type
Article
Abstract
BACKGROUND: The Pipeline Flex embolization device has several advantages over the first-generation Pipeline embolization device (Covidien, Dublin, Republic of Ireland). Despite these advantages, we have observed frequent difficulty in recapturing the device's delivery wire. OBJECTIVE: To prospectively document the incidence of failure to recapture the delivery wire. METHODS: We tracked our experience in patients undergoing endovascular treatment with a flow-diverting stent for cerebral aneurysms between 1 May and 30 September 2016. Patient and lesion characteristics, device dimensions, and technical outcomes of delivery wire recapture were prospectively recorded for each device. RESULTS: Eighteen devices were deployed in 15 patients by the senior author (FCA) during this period. Failure to recapture the delivery wire occurred in 10 of 18 (56%) cases. No adverse outcomes of delivery wire recapture failure were encountered in this series. CONCLUSIONS: The incidence of delivery wire recapture failure with the Pipeline Flex device is high. Failure to recapture the delivery wire carries a theoretical risk of stent displacement when re-navigating across the device, and endovascular surgeons should be aware of this limitation. Authors are encouraged to report delivery wire recapture failure rates in future clinical series in which the Pipeline Flex device is used.
Keywords
Aneurysm, Flow Diverter, Stent, guidewire
Medical Subject Headings
Adult; Aged; Embolization, Therapeutic (instrumentation, methods); Endovascular Procedures (methods); Female; Humans; Incidence; Intracranial Aneurysm (diagnostic imaging, therapy); Male; Middle Aged; Prospective Studies; Stents; Treatment Failure
Publication Date
6-1-2017
Publication Title
Journal of neurointerventional surgery
E-ISSN
1759-8486
Volume
9
Issue
6
First Page
571
Last Page
573
PubMed ID
28100649
Digital Object Identifier (DOI)
10.1136/neurintsurg-2016-012856
Recommended Citation
Mooney, Michael A.; Moon, Karam; Gross, Bradley A.; Ducruet, Andrew F.; and Albuquerque, Felipe C., "Incidence of delivery wire recapture failure with the Pipeline Flex device" (2017). Translational Neuroscience. 2109.
https://scholar.barrowneuro.org/neurobiology/2109