In Vitro Biomechanical Evaluation Of A Novel Minimally Invasive Sacroiliac Joint Fixation Device

Department

neurobiology

Document Type

Article

Abstract

Background: Sacroiliac (SI) joint pathology may result in low-back pain, which causes substantial disability. Treatment failure with operative management of SI pain may be related to incomplete fusion of the joint and to fixation failure. The objective of this study was to evaluate the initial biomechanical stability of SI joint fixation with a novel implantable device in an in vitro human cadaveric model. Methods: The right and left sides of 3 cadaveric L4-pelvis specimens were tested (1) intact, (2) destabilized, and (3) instrumented with an implantable SI joint fixation device using a simulated single-stance load condition. Right-leg and left-leg stance data were grouped together for a sample size of 6, and angular range of motion (ROM) was determined during application of flexion-extension, lateral bending, and axial rotation bending moments to a limit of 7.5 Nm. Results: Following intact testing, destabilization by severing the posterior SI joint capsule and ligaments and the pubic symphysis reliably produced a significantly destabilized joint with the mean angular ROM more than doubling in flexion-extension and lateral bending and more than tripling in axial rotation (P .003) compared to the intact condition. Instrumentation with the SI screw fixation device significantly reduced mean joint ROM compared to the destabilized condition in all 3 anatomic planes tested (P, .001). When compared to the intact condition, the SI-instrumented condition significantly reduced lateral bending (P ¼ .01) and had a similar ROM in flexion-extension (P ¼ .14) and axial rotation (P ¼ .85). Conclusions: Instrumentation with the SI screw fixation device significantly reduced mean joint ROM compared to the destabilized condition, with similar ROM in flexion-extension and axial rotation, and it significantly reduced ROM in lateral bending compared to that for the intact joint. The ROM values observed with the instrumented condition were comparable to levels of mobility considered favorable for spinal fusion.

Publication Date

10-1-2018

Publication Title

International Journal of Spine Surgery

ISSN

22114599

Volume

12

Issue

5

First Page

603

Last Page

610

Digital Object Identifier (DOI)

10.14444/5072

This document is currently not available here.

Share

COinS