Biomechanical Analysis Of An Expandable Lumbar Interbody Spacer
Department
neurobiology
Document Type
Article
Abstract
Objective: Recently developed expandable interbody spacers are widely accepted in spinal surgery; however, the resulting biomechanical effects of their use have not yet been fully studied. We analyzed the biomechanical effects of an expandable polyetheretherketone interbody spacer inserted through a bilateral posterior approach with and without different modalities of posterior augmentation. Methods: Biomechanical nondestructive flexibility testing was performed in 7 human cadaveric lumbar (L2–L5) specimens followed by axial compressive loading. Each specimen was tested under 6 conditions: 1) intact, 2) bilateral L3–L4 cortical screw/rod (CSR) alone, 3) WaveD alone, 4) WaveD + CSR, 5) WaveD + bilateral L3–L4 pedicle screw/rod (PSR), and 6) WaveD + CSR/PSR, where CSR/PSR was a hybrid construct comprising bilateral cortical-level L3 and pedicle-level L4 screws interconnected by rods. Results: The range of motion (ROM) with the interbody spacer alone decreased significantly compared with the intact condition during flexion–extension (P = 0.02) but not during lateral bending or axial rotation (P ≥ 0.19). The addition of CSR or PSR to the interbody spacer alone condition significantly decreased the ROM compared with the interbody spacer alone (P ≤ 0.002); and WaveD + CSR, WaveD + PSR, and WaveD + CSR/PSR (hybrid) (P ≥ 0.29) did not differ. The axial compressive stiffness (resistance to change in foraminal height during compressive loading) with the interbody spacer alone did not differ from the intact condition (P = 0.96), whereas WaveD + posterior instrumentation significantly increased compressive stiffness compared with the intact condition and the interbody spacer alone (P ≤ 0.001). Conclusions: The WaveD alone significantly reduced ROM during flexion–extension while maintaining the axial compressive stiffness. CSR, PSR, and CSR/PSR hybrid constructs were all effective in augmenting the expandable interbody spacer system and improving its stability.
Publication Date
6-1-2018
Publication Title
World Neurosurgery
ISSN
18788750
Volume
114
First Page
e616
Last Page
e623
Digital Object Identifier (DOI)
10.1016/j.wneu.2018.03.041
Recommended Citation
Soriano-Baron, Hector; Newcomb, Anna G.U.S.; Malhotra, Devika; Palma, Atilio E.; Martinez-del-Campo, Eduardo; Crawford, Neil R.; Theodore, Nicholas; Kelly, Brian P.; and Kaibara, Taro, "Biomechanical Analysis Of An Expandable Lumbar Interbody Spacer" (2018). Translational Neuroscience. 104.
https://scholar.barrowneuro.org/neurobiology/104