Effects of minimum stimulation settings for the Med El Tempo+ speech processor on speech understanding
Document Type
Article
Abstract
OBJECTIVE: The aim of this study was to assess the effects of variations in the settings for minimum stimulation levels on speech understanding for adult cochlear implant recipients using the Med El Tempo+ speech processor. DESIGN: Fifteen patients served as listeners. The test material included sentences presented at a conversational level in noise (74 dB SPL at +10 dB signal-to-noise ratio), sentences presented at a soft level in a quiet background (54 dB SPL), consonants in "vCv" environment (74 dB SPL re: vowel peaks), and synthetic vowels in "bVt" environment (54 dB SPL re: vowel peaks). The patients' speech processors were programmed with minimum stimulation levels set to behavioral threshold, set to 10% of most comfortable loudness, and set to 0 muA. RESULTS: The level of speech understanding achieved in the behavioral threshold condition was not significantly different from that achieved in either the 10% of most comfortable loudness or 0 muA conditions for any test material. Only 2 of the 15 patients demonstrated performance differences of greater than 10 percentage points between the 0 muA condition and the behavioral threshold condition on more than a single test. CONCLUSIONS: Our results demonstrate that there are no grievous consequences, in terms of speech understanding, for setting minimum stimulation levels below behavioral thresholds. The time savings from setting thresholds to 10% of MCL or 0 muA may be especially useful during the initial device fitting.
Medical Subject Headings
Acoustic Stimulation; Adult; Analysis of Variance; Auditory Threshold; Cochlear Implants; Female; Hearing Loss (therapy); Humans; Male; Prosthesis Fitting; Speech Discrimination Tests; Speech Perception; Treatment Outcome
Publication Date
8-1-2005
Publication Title
Ear and hearing
ISSN
0196-0202
Volume
26
Issue
4 Suppl
First Page
2S
Last Page
6S
PubMed ID
16082262
Digital Object Identifier (DOI)
10.1097/00003446-200508001-00002
Recommended Citation
Spahr, Anthony J. and Dorman, Michael F., "Effects of minimum stimulation settings for the Med El Tempo+ speech processor on speech understanding" (2005). ENT and Skull Base Surgery. 136.
https://scholar.barrowneuro.org/ent-and-skull-base-surgery/136