Impact of multipoint pelvic fixation and multirod distal constructs on proximal junction biomechanics in cadaveric specimens
Document Type
Article
Abstract
OBJECTIVE: Multipoint pelvic fixation with multirod constructs is increasingly used for long-segment deformity constructs to reduce rates of distal failure. However, more robust distal fixation may negatively impact proximal junction biomechanics, and this potential relationship has not been extensively studied. METHODS: Standard nondestructive flexibility tests (7.5 Nm) were performed on 7 cadaveric specimens (L1-pelvis) to assess intervertebral flexibility (range of motion [ROM]), rod strain, and screw bending moments along a posterior fusion construct (pedicle screw and rod [PSR]) spanning L2-S1, supplemented by bilateral primary S2 alar-iliac (S2AI) fixation (2 S2AI screws and 2 rods), followed by additional S2AI screw placement and bilateral accessory rod placement spanning L4-S2AI (4 S2AI screws and 4 rods). Four conditions were tested for each specimen: 1) intact; 2) L2-S1 PSR; 3) L2-S2AI PSR; and 4) L2-S2AI plus L4-S2AI. Data were analyzed using repeated-measures ANOVA. RESULTS: Seven cadaveric specimens were included. Proximal rod strain at the L2-3 level did not change across the varying test conditions in the 7 specimens tested (p > 0.05 for all conditions). There was no significant difference detected in proximal screw strain across conditions (p > 0.05). Finally, no significant difference was found in L2-3 ROM (p > 0.05) across instrumented variations, all of which were more rigid than intact specimens. CONCLUSIONS: Pelvic fixation with 2 or 4 screws and 2 or 4 rods, respectively, did not significantly alter proximal junction screw or rod strain in a cadaveric model. Robust pelvic fixation might protect against distal failure without deleterious effects on the proximal junction.
Keywords
accessory rods, multipoint pelvic fixation, proximal junctional kyphosis, sacral
Medical Subject Headings
Humans; Biomechanical Phenomena (physiology); Cadaver; Spinal Fusion (methods, instrumentation); Range of Motion, Articular (physiology); Lumbar Vertebrae (surgery); Pelvis (surgery); Bone Screws; Sacrum (surgery); Aged; Pelvic Bones (surgery); Male; Female; Pedicle Screws; Middle Aged
Publication Date
11-1-2025
Publication Title
Journal of neurosurgery. Spine
E-ISSN
1547-5646
Volume
43
Issue
5
First Page
540
Last Page
546
PubMed ID
40882238
Digital Object Identifier (DOI)
10.3171/2025.4.SPINE25263
Recommended Citation
Rudy, Robert F.; Sawa, Anna G.; McBryan, Sarah; Mugge, Luke A.; Thielen, Katherine; Assefa, Temesgen G.; Lindsey, Derek P.; Polly, David W.; Uribe, Juan S.; Kelly, Brian P.; and Turner, Jay D., "Impact of multipoint pelvic fixation and multirod distal constructs on proximal junction biomechanics in cadaveric specimens" (2025). Translational Neuroscience. 2467.